The FREE System is an integrated analytical system
consisting of a photometer and a mini-centrifuge and a number of
tools (microipipettes, tips, tubes etc.), that is able to carry
out any chemical analyses based on the principles of spectrophotometry
(e. g. blood glucose, cholesterol, creatinine, transaminases etc)
and, particularly, the whole panel of oxidative stress tests.
TECHNICAL CHARACTERISTICS
Photometric system
Light source: 20 W long life halogen bulb.
Spectral range: 320 -680 nm.
Filters: 340, 405, 505, 546, 578, 630, with 8 nm bandwidth.
Placing for optional filters: two
Photo detector: solid state.
Termostating
Dry, with Peltier cooling -heating element.
Temperature: 20° - 45°C.
Sensibility: 0,1°C.
Stabilizing time: from 25° to 37° in 7 minutes.
Thermostatic block: 9 places for square and cylindrical cuvettes.
Measuring system
Zero setting: automatic.
Measuring range: -200 +2500 in OD
Photometric linearity: better than 1%.
Photometric accuracy: ± 2% at 700 00
Repeatability: ±1 digit.
Drift: lower than .005 in 0.0 per hour.
Measuring time: 0,3 seconds
Data programming and data output
Keyboard: 12 keys for entering alphanumerical data, 2 print
command keys, 3 photometer running keys, 1 key for
timer.
Display: Liquid crystal alphanumeric, with 4 lines of 20
characters.
Printer : Graphic, 192 points per line, for thermosensitive paper.
General Characteristics
~
Power supply : 115 -230 VAC
50-60Hz
Consumption : 83 VA
Fuse line : T 1A L
Operating conditions
Temperature : 15 - 32 °C (operating)
0 - 50 °C (non operating)
Relative humidity: 20 - 80 % (operating)
0 - 90 % (non operating)
Altitude: < 2000 m (operating)
Safety : DIRECTIVE 73/23/EEC
REGULATIONS: CEI-EN 61010-1, CLASSE I,
INSTALLATION CATEGORY II
Electromagnetic DIRECTIVE 89/336/CEE
compatibility:
Dimensions: 30 x 30 x 36 (h) cm
Weight : 8 Kg
|